| Model Number |
7121/60 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented in clinic for a follow up.The patient's right ventricular (rv) lead exhibited noise over-sensing.Lead revision was discussed, but no interventions were made or reported.The patient was in stable condition.
|
| |
|
Event or Problem Description
|
|
New information noted that the right ventricular lead kept oversensing noise.The patient had presented for a generator change on august 5, 2022, but the lead revision was not performed due to patient access issue.The lead remained active and would be monitored.The patient was stable.
|
| |
|
Event or Problem Description
|
|
New information noted that the right ventricular lead was capped and replaced (b)(6) 2022.The patient was in stable condition throughout the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of oversensing was not confirmed.As received, only the connector portion of the lead was returned in one piece.Electrical, visual, and x-ray examination of the lead did not find any anomalies with the exception of procedural damage.
|
| |
|
Search Alerts/Recalls
|