|
|
| Model Number |
SC45A |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Sc45a - jaw stuck.It can't be open.Sc45a - jaw stuck.It can't be open.Was surgery delayed due to the reported event? unknown, was procedure successfully completed? unknown, were fragments generated? unknown, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study: unknown, (b)(4).Device property of: none, device in possession of: none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # 444a43.Date of event is 2022, event day and month unknown.Captured as 1/1/2022.Additional information was requested, and the following was obtained: 1.Was the device locked on tissue with the jaws closed? no.If yes, how was the device removed? 2.Did the jaws eventually open? no.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that one sc45a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties noted during the functional testing, the device opened and closed as expected.The staple line and cut line were complete and the staples meet the staple release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: if the jaws do not automatically open after the anvil release button is pressed, first ensure that the knife is in the retracted position by verifying that the stroke count indicator displays ¿0¿ and knife direction indicator points towards the proximal side of the instrument, or that the knife blade indicator is in the home position.If the stroke count indicator or knife blade indicator is not in the home position, push the red manual knife reverse switch downward to reverse the knife motion and squeeze the firing trigger, completely until it rests on the closing trigger.Press the anvil release button.If the jaws do not open at this point, then gently pull the closing trigger, upward (away from the handle) until both firing and closing triggers return to their original positions.The event could not be confirmed as the device performed as expected and during the functional testing, the device opened and closed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|