Investigation: visual inspection: the following observations were made during the visual inspection: deposits on the pediatric prechamber, deposits in the inlet catheter, ruptured distal catheter.Measurement of surface of the housing: deformation detected: yes - measured value: -0,05 mm [tolerance (0 ± 0.02mm)] too much pressure on the valve, for example through the adjustment tool or resulting from a fall or hit to the head of the patient, can compromise the integrity of the valve.Permeability test: the test showed that the progav shunt system is permeable.Additionally, we tested the pediatric prechamber.Out test determined that the prechamber is permeable.During the tests, some particles were flushed out.Computer control test: the computer controlled test showed the opening pressure of the progav, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 12.62 cmh2o.This is within the specified tolerance of 11 cmh2o ± 3 cmh2o.Additionally, the shuntassistant was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 20 cmh2o in the vertical position, a pressure of 20 cmh2o +4/-2 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shuntassistant® had a pressure of 19.21 cmh2o.This is within the specified tolerance of 20 cmh2o +4/-2 cmh2o.Adjustability test: the progav was found to be adjustable to all pressure settings.The shuntassistant is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force of the progav was within the specified tolerance and the brake function operated as expected.Internal inspection of product: in order to verify whether the investigated shunt system was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valves are finally opened with a special tool.After dismantling of the valves, deposits were found in both valves.Results: based on our investigation results, we can determine a deformation of the outer housing of progav, a rupture of the distal catheter and deposits in both valves.The deposits had no affect to the technical properties at the time of the investigation.The cause of the deformation and the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.It might be a result of growth that the catheter has been torn off.Nevertheless, it is retrospective not possible for us to determine the exact cause of the rupture.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
|