| Model Number |
RE9CIC-W-MP |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).Device was received for remake on 07-05-2022.Remake was performed and shipped on 07-11-2022.Event investigation is still in progress.Please notice no further investigation of the device will be performed, only the investigation related to the event will be performed.Dhr has been reviewed.Based on the available information, the case indicates need for reporting.User's right ear becomes inflamed every 6-9 month, and requires a doctor visit and ear drops (unknown).Based on the information provided the event did lead to or could potential lead to a serious injury.Final report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that 41 year old male patient has infections in his ear due to hearing aids so they want to try remaking it so that it fits better and is not irritating the inside of the ear.Every 6 to 9 months, user's right ear canal would become inflamed and require a doctor's visit and a week of ear drops (unknown) in order for the device to fit again.The user experienced redness, pain, and swelling.Since the user had the device almost 3 years ago the right ear has become inflamed 3 times.Each time required a visit to the doctor and prescription of ear drops until swelling and redness went away.Recommendation from hcp was to remake the device.The patient has maintained(cleaned) the device as specified in the user guide.The reaction is related to the physical fit of the device and a remake of the device has been performed.The patient has no history of allergies, hypersensitivity or fungal infections.The patient had not used another hearing aid before and experienced similar events.The medical diagnosis and other relevant medical history is unknown.Onset and stop date of the event is unknown.Hcp unsure of start and stop date, the patient had to use drops (unknown details) for a few weeks before it cleared up.Reports the patient is recovered.
|
| |
|
Event or Problem Description
|
|
It was reported that 41 year old male patient has infections in his ear due to hearing aids so they want to try remaking it so that tit fits better and is not irritating the inside of the ear.Every 6 to 9 months, user's right ear canal would become inflamed and require a doctor's visit and a week of ear drops (unknown) in order for the device to fit again.The user experienced redness, pain, and swelling.Since the user had the device almost 3 years ago the right ear has become inflamed 3 times.Each time required a visit to the doctor and prescription of ear drops until swelling and redness went away.Recommendation from hcp was to remake the device.The patient has maintained(cleaned) the device as specified in the user guide.The reaction is related to the physical fit of the device and a remake of the device has been performed.The patient has no history of allergies, hypersensitivity or fungal infections.The patient had not used another hearing aid before and experienced similar events.The medical diagnosis and other relevant medical history is unknown.Onset and stop date of the event is unknown.Hcp unsure of start and stop date, the patient had to use drops (unknown details) for a few weeks before it cleared up.Reports the patient is recovered.Follow up received on 5-aug-2022.The doctor prescribed following medication: cprofloxacin 0.3% and dexamethasone 0.1%.Ndc: 0781-6186-67.Manufacturer: sandoz.4 drop a day.Consider this final report.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref#: (b)(4).Device was received for remake on 07-05-2022.Remake was performed and shipped on 07-11-2022.Event investigation is still in progress.Please notice no further investigation of the device will be performed, only the investigation related to the event will be performed.Dhr has been reviewed.Based on the available information, the case indicates need for reporting.User's right ear becomes inflamed every 6-9 month, and requires a doctor visit and ear drops (unknown).Based on the information provided the event did lead to or could potential lead to a serious injury.Final report will be submitted upon completion of the investigation.25-aug-2022 technical investigation (dhr) concluded: no deviation or changes were found during manufacturing of the device.The clinical investigation concluded: clinical conclusion on the case is that it is likely that the reaction is a reaction to the shell material or to fit and changing shell material or remaking the device as hcp has requested is appropriate.Allergic reaction is identified in the risk analysis and mitigated to an acceptable level through device design by compliance with standards for biocompatibility.Risk register reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register.Corrections and additional information: b5: final follow up information added.E1: initital reporter information corrected and updated.H6: codes 'investigation findings' and ''investigation conlusion' corrected.Consider this final report.
|
| |
|