Model Number CQ75124
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Five photos were provided and reviewed.The first photo shows an inflation device connected to the inflation hub of the catheter and a guidewire inserted in the guidewire hub of the catheter.Liquid is seen in the inflation device.The second photo shows the conquest balloon loaded onto a guidewire.The third photo shows a compound rupture noted on the balloon.The fourth photo shows a compound rupture noted on the balloon.The fifth photo shows a compound rupture noted on the balloon.Therefore, the investigation is confirmed for the reported compound balloon rupture as a compound rupture was noted on the balloon from the photo review.A definitive root cause for the reported compound balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 07/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured at 9 atm.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15119314
Report Number2020394-2022-00615
Device Sequence Number20649896
Product Code DQY
UDI-Device Identifier00801741063251
UDI-Public(01)00801741063251
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ75124
Device Catalogue NumberCQ75124
Device Lot NumberREFV3476
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2022
Initial Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-