Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Five photos were provided and reviewed.The first photo shows an inflation device connected to the inflation hub of the catheter and a guidewire inserted in the guidewire hub of the catheter.Liquid is seen in the inflation device.The second photo shows the conquest balloon loaded onto a guidewire.The third photo shows a compound rupture noted on the balloon.The fourth photo shows a compound rupture noted on the balloon.The fifth photo shows a compound rupture noted on the balloon.Therefore, the investigation is confirmed for the reported compound balloon rupture as a compound rupture was noted on the balloon from the photo review.A definitive root cause for the reported compound balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 07/2024).
|