• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: I-SENS, INC. ASSURE PRISM MULTI BLOOD GLUCOSE TEST SYSTEM; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

I-SENS, INC. ASSURE PRISM MULTI BLOOD GLUCOSE TEST SYSTEM; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number GM505RA
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the complaint products were not returned and sn or lot was not identified, i-sens analyzed the precision issue with a reference meter of the same model and reference strips.As a result of the control solution test, all results were within the standard range, and the cv% was within the acceptance criteria (below 5%), which satisfied the internal standard in i-sens.Therefore, it is judged that the product operated normally.
 
Event or Problem Description
Rep called customer after receiving an email from sales rep about issue.Asked customer if she has all three meters that had issues.Customer does not have the meters.Customer threw one away thinking it was just a one-off issue.Rep went over some variables that can cause the fluctuations in readings.Rep asked customer how many meters she has, and customer stated she has one she can get a serial number from, and she would have to call the station to get the other information.First issue was last thursday or friday with the meter she has.There was about a 200-point difference.Ems was dispatched for a diabetic emergency.Patient wasn't feeling well.Fire was on the scene first, assessed the patient, took blood sugar and reading was 110.When the ambulance got there, the patient's blood sugar was 300.Readings were within 10 minutes.No medications were administered to the patient, all they did was put a saline line in and did 8-10 ccs to flush it.Verified it was a fingerstick test and they disinfected finger before taking the reading.Rep asked about storage and combining strips and caller indicated they store in the same bottle they come in.Customer ended call to take another call.Rep.Called back later to complete call.Customer was unable to get the serial number of the meter like she thought.Rep provided his email, phone number and extension for customer to provide serial number later.Meter and strips are stored together in a jump bag in the side panel of the truck.Rep explained temperature fluctuations can cause issues.Customer will call back in a few days with the other meter having issues.Replaced product and sent return label.On (b)(6) 2020, rep followed up to get information and had to leave a voicemail.Caller never sent product details, did not return meter, and has not called back since.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURE PRISM MULTI BLOOD GLUCOSE TEST SYSTEM
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
I-SENS, INC.
43, banpo-daero, 28-gil
seocho-gu
seoul, 06646
KS  06646
Manufacturer (Section G)
I-SENS WONJU FACTORY
94-1, donghwagongdan-ro
munmak-eup
wonju-si, 26365
KS   26365
Manufacturer Contact
hee seon
43, banpo-daero, 28-gil
seocho-gu
seoul, 06646
MDR Report Key15119584
Report Number3005621222-2022-00007
Device Sequence Number7286438
Product Code NBW
UDI-Device Identifier00015482530013
UDI-Public(01)00015482530013
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberGM505RA
Device Catalogue Number530001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2022
Initial Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-