• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60WF
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.No other complaint reported in the lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during the intraocular lens implant procedure, it was found that lens has mark on it.The lens has been replaced with the same model lens.There was no patient harm reported.Additional information has been requested.
 
Additional Manufacturer Narrative
The lens was returned in the lens case.Viscoelastic was observed on the lens.The lens had been cut into two pieces across the center of the optic.One portion of the optic had an area that had the appearance of lens damage.The area was swabbed.A clear linear material was removed from the optic surface.The material was isolated between glass slides and sent to the lab for identification testing.The clear material was isolated and analyzed using microscopic fourier transform infrared spectroscopy (micro ft-ir).Comparison of the opaque residues generated spectra to a library of spectra found the best match to be company cartridge base coat.It cannot be determined if the observed material may have been interpreted as the reported scratch or if the scratch was obscured by the cut across the center of the optic.A qualified iol and handpiece were indicated.The viscoelastic used was not provided.It is unknown if qualified product was used.The product investigation could not identify a root cause for the reported complaint.It cannot be determined if the observed material may have been interpreted as the reported scratch or if the scratch was obscured by the cut across the center of the optic.Based on the iol evaluation and the lab testing results, the reported scratch may have been the observed material.The clear material was isolated and analyzed using microscopic fourier transform infrared spectroscopy (micro ft-ir).Comparison of the opaque residues generated spectra to a library of spectra found the best match to be company cartridge base coat.It is unknown if a qualified viscoelastic was used.Per the instruction for use (ifu): the company iol delivery system is for implantation of qualified company foldable iols.No unqualified lenses should be used with the company iol delivery system.The company cartridges are qualified for use with compatible company handpieces for the surgical implantation of company qualified foldable iols.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15119637
Report Number1119421-2022-01648
Device Sequence Number5614747
Product Code HQL
UDI-Device Identifier00380652332729
UDI-Public00380652332729
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/08/2026
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.230
Device Lot Number15145527
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer10/11/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH CARTRIDGE C; MONARCH CARTRIDGE D; MONARCH INJECTOR