| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Foreign source: japan.The product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 01725, 0001825034 - 2022 - 01727.
|
| |
|
Event or Problem Description
|
|
It was reported that during incoming inspection, the sterile packaging was found to be damaged.It is unknown if the sterility of the device has been compromised.There was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned product identified damage to the sterile packaging (blister).Sterility has not been compromised.Reported event has been confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Search Alerts/Recalls
|