MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Medical Device Problem Codes Thermal Decomposition of Device (1071); Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The product has been returned to zimmer biomet, and the investigation is in process.Once the investigation is completed, a supplemental medwatch 3500a will be submit h3 other text : investigation incomplete.
 
Event or Problem Description
It was reported that the power inlet module was burnt.The event timing was not during surgery.No patient involvement or harm.It was reported that no further information is available.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.The results of the investigation are as follows: review of the most recent repair record determined that the cart was not powering on.The power inlet module was replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH POWER IV
Common Device Name
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15121029
Report Number0001954182-2022-00069
Device Sequence Number15284664
Product Code JCX
UDI-Device Identifier00889024501621
UDI-Public(01)00889024501621
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K190789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00514010300
Device Lot Number0042550
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2022
Supplement Date Received by Manufacturer08/29/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/11/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-