MEDTRONIC PUERTO RICO OPERATIONS CO. 670G INSULIN PUMP MMT-1780KL; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-1780KL
Medical Device Problem Codes Break (1069); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Hyperglycemia (1905); Hypoglycemia (1912)
Date of Event 03/26/2020
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information that damage (physical or cosmetic) occurred.There was no adverse impact or consequence reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).Pump passed displacement test.Unit was received with cracked reservoir tube retainer.The p-cap locks properly into place.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
670G INSULIN PUMP MMT-1780KL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key15121198
Report Number2032227-2022-304184
Device Sequence Number5646418
Product Code OZP
UDI-Device Identifier00763000166526
UDI-Public(01)00763000166526(17)220708
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/08/2022
Device Model NumberMMT-1780KL
Device Catalogue NumberMMT-1780KL
Device Lot NumberHG3KQQ1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2020
Initial Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/04/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
Patient Weight87 KG
-
-