• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EIT EMERGING IMPLANT TECHNOLOGIES GMBH POST IBF UI H 10MM 4DEG 26/9; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EIT EMERGING IMPLANT TECHNOLOGIES GMBH POST IBF UI H 10MM 4DEG 26/9; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number PUI41006
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2022, patient underwent a tlif spinal procedure (date unknown) and some point afterward the tlif spacer extruded posteriorly into the spinal canal.Patient returned to the surgeon complaining of left leg pain.Upon x-ray, the mal-positioned spacer was discovered and a revision surgery was scheduled.On (b)(6) 2022 patient underwent an alif procedure to remove the tlif spacer and replace with a synfix alif stand-alone spacer.This report is for one (1) post ibf ui h 10 mm 4deg 26/9.This is report 1 of 1 for complaint: (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the photo revealed that the post ibf ui h 10 mm 4deg 26/9 moved posteriorly from where it was previously implanted.The reported event of migration can be confirmed with the evidence provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for post ibf ui h 10 mm 4deg 26/9, pui41006 there is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: part # pui41006, lot # e21la0755, release to warehouse date: 08 oct 2021 supplier: (b)(4).No ncr's were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POST IBF UI H 10MM 4DEG 26/9
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
EIT EMERGING IMPLANT TECHNOLOGIES GMBH
eisenbahnstrasse 84
wurmlingen D-785 73
DE  D-78573
Manufacturer (Section G)
EIT EMERGING IMPLANT TECHNOLOGIES GMBH
DE  
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15121201
Report Number3013730328-2022-00063
Device Sequence Number14864442
Product Code MAX
UDI-Device Identifier04260557778517
UDI-Public(01)04260557778517
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170503
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPUI41006
Device Catalogue NumberPUI41006
Device Lot NumberE21LA0755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2022
Initial Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/08/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
-
-