• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTAWIRE AND WIRECLIP TORQUER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTAWIRE AND WIRECLIP TORQUER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 29710
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that there was blood-tinged tissue present on the wire.The 100% stenosed target lesion was located in the left superficial femoral artery.A peripheral rotawire guidewire was selected for use.During the procedure, a 2.00mm peripheral rotalink plus system began making a high-pitched noise and the burr seemed to be stuck as the physician attempted to move it slightly back and forth without success.The physician then switched to dynaglide mode to completely remove the burr out of the patient's body and was able to do so.When the rotawire was removed from the patient's body, it was noted that there was blood-tinged tissue present on the wire.The procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for analysis.Visual and microscopic inspection revealed that the device has a kink in the body of the guidewire located approximately at 45 cm from the proximal end.Residue was noted stuck in the body of the guidewire.Therefore, the reported event was confirmed.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that there was blood-tinged tissue present on the wire.The 100% stenosed target lesion was located in the left superficial femoral artery.A peripheral rotawire guidewire was selected for use.During the procedure, a 2.00mm peripheral rotalink plus system began making a high-pitched noise and the burr seemed to be stuck as the physician attempted to move it slightly back and forth without success.The physician then switched to dynaglide mode to completely remove the burr out of the patient's body and was able to do so.When the rotawire was removed from the patient's body, it was noted that there was blood-tinged tissue present on the wire.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERIPHERAL ROTAWIRE AND WIRECLIP TORQUER
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15121456
Report Number2134265-2022-08205
Device Sequence Number5614807
Product Code MCW
UDI-Device Identifier08714729863649
UDI-Public08714729863649
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/22/2024
Device Model Number29710
Device Catalogue Number29710
Device Lot Number0029094950
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer08/12/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age82 YR
Patient SexFemale
Patient Weight56 KG
Patient RaceWhite
-
-