| Model Number |
MAJ-411 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Only one (1) distal cover malfunctioned.The facility did not know which of the four (4) distal covers at the facility malfunctioned.Therefore, four (4) distal covers would be sent for evaluation.The device was returned to an olympus service center for evaluation and the reported issue was not confirmed.The distal cover satisfied the specified value of fixing strength.There was no problem with the locking mechanism of the caps.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus, after use and during bedside cleaning, the distal cover came off just by gently stroking the tip of the evis lucera duodenovideoscope.The customer stated the distal cover may come off too easy.No patient injury reported.It was unclear whether the issue was due to the distal cover or scope.This complaint is related to patient identifiers; (b)(6)(evis lucera duodenovideoscope, jf-260v), (b)(6) (distal cover, maj-411), (b)(6) (distal cover, maj-411), (b)(6) (distal cover, maj-411).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the approved final investigation.The device history record was unable to be reviewed for this device since the serial number was unknown.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, and since no device malfunction was confirmed during evaluation, the definitive root cause of the distal cover detachment could not be determined.It is possible that imperfection of the attachment method of the cover may have led to the unfixing of the cover to the hook at the distal end of the scope, which did not engage the detachment prevention lock.The maj-311/maj 411 abridged instruction manual identifies verbiage which may have prevented the phenomenon.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the approved final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the definitive root cause of the distal cover issue could not be determined.The lock mechanism of the cover did not have any problem.It is possible that there was an imperfection of the attachment method of the cover, which may have led to unfixing of the cover to the hook at the distal end of the scope, and the lock of the detachment prevention was not engaged.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|