• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. DISTAL COVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. DISTAL COVER Back to Search Results
Model Number MAJ-411
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Only one (1) distal cover malfunctioned.The facility did not know which of the four (4) distal covers at the facility malfunctioned.Therefore, four (4) distal covers would be sent for evaluation.The device was returned to an olympus service center for evaluation and the reported issue was not confirmed.The distal cover satisfied the specified value of fixing strength.There was no problem with the locking mechanism of the cap.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
The customer reported to olympus, after use and during bedside cleaning, the distal cover came off just by gently stroking the tip of the evis lucera duodenovideoscope.The customer stated the distal cover may come off too easy.No patient injury reported.It was unclear whether the issue was due to the distal cover or scope.This complaint is related to patient identifiers; (b)(4) (evis lucera duodenovideoscope, jf-260v), (b)(4) (distal cover, maj-411), (b)(4) (distal cover, maj-411), (b)(4) (distal cover, maj-411).
 
Additional Manufacturer Narrative
This complaint is related to patient identifiers; (b)(6) (evis lucera duodenovideoscope, jf-260v), (b)(6) (distal cover, maj-411), (b)(6) (distal cover, maj-411) and (b)(6) (distal cover, maj-411).This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial/lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the distal cover fixing value and cover lock mechanism were found to meet specification, no problems were found.The root cause of the issue could not be determined, however, it is likely that the distal cover detachment prevention lock was not properly engaged, resulting in the detachment issue.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTAL COVER
Common Device Name
DISTAL COVER
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15122128
Report Number9610595-2022-00374
Device Sequence Number7286593
Product Code FET
UDI-Device Identifier04953170071416
UDI-Public04953170071416
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K954451
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-411
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2022
Supplement Date Received by Manufacturer03/27/2023
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date04/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-