Model Number QUANTUM Q6 EDGE HD
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Bruise/Contusion (1754); Laceration(s) (1946)
Date of Event 01/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
 
Event or Problem Description
Consumer alleges he was asleep in his chair and in the middle of the night the back allegedly broke away and he allegedly fell to the floor.
 
Additional Manufacturer Narrative
The right side seat base plate interface was torn/broken.The left side back to base slot block (mec121721), spacer and associated hardware were missing.With this, it is likely the seat interface was compromised on the opposing attached side due to these critical missing components.The device history record indicates the seating system was built properly in 2018.Although it is unknown why these parts were missing, visual inspection indicates adjustments to the seat/back depth were made post final release suggesting the interface components were loosened during these adjustments.Further, the seating system was subjected to heavy incontinence.It is unknown if this played any roll in the failure.
 
Event or Problem Description
Consumer alleges he was asleep in his chair and in the middle of the night the back allegedly broke away and he allegedly fell to the floor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
POWERED WHEELCHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york ave
duryea, PA 18642
8008008586
MDR Report Key15122909
Report Number2530130-2022-00067
Device Sequence Number7267690
Product Code ITI
UDI-Device Identifier00606509200059
UDI-Public00606509200059
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K042612
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberQUANTUM Q6 EDGE HD
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2022
Supplement Date Received by Manufacturer07/22/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-