U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); Visual Disturbances (2140); Halo (2227); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site as it was discarded; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints revealed 1 other complaint has been received for this production order number and it was voided as it was a duplicate serial number.No escalation was required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.Attempts were made to obtain the missing information; however, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was explanted in a secondary surgical procedure due to the patient experiencing unbearable halos, shadowing, vertical doubling, and was unable to function with the lens.Pre-operative visual acuity (va) was 20/40 -1 and post-operative va was 20/60-2, vision corrected to 20/20.The patient had posterior capsule opacification (pco), but opted not to have yag performed but to explant instead.No incision enlargement, sutures, or vitrectomy were required.The device was discarded.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15122961
Report Number3012236936-2022-01955
Device Sequence Number12196668
Product Code HQL
UDI-Device Identifier05050474750647
UDI-Public(01)05050474750647(17)240808
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/08/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Initial Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-