• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ENTERAL FEEDING PUMP; ENTERAL INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ENTERAL FEEDING PUMP; ENTERAL INFUSION PUMP Back to Search Results
Model Number INFKIT2
Device Problem Use of Device Problem (1670)
Patient Problem Aspiration/Inhalation (1725)
Event Date 06/29/2022
Event Type  Death  
Event Description
The initial reporter states that the patient passed away during the night.They stated that the patient used the pump at night for feedings at a rate of 55 ml/hr and a continuous dose.They suspected that the patient had aspirated and died.Mmdg did follow up with the initial reporter, but they did not provide any additional information about the event or the patients official cause of death.(b)(4).
 
Manufacturer Narrative
The device was returned to mmdg for evaluation.A dhr was completed and found no non-conformances.When the device was returned to mmdg for investigation, it operated as expected.There is no indication that the device malfunctioned.Based on the information provided to mmdg and the investigation, the pump appears to have performed as expected.This report is being filed out of an abundance of caution due to the patients death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ENTERAL FEEDING PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin egelhoff
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key15123069
MDR Text Key296793586
Report Number1722139-2022-00739
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberINFKIT2
Device Catalogue NumberINFKIT2
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2022
Date Manufacturer Received06/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age2 YR
Patient SexFemale
-
-