• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. NON-CF RESECTOSCOPE METAL SHEATH, 24 FR; ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. NON-CF RESECTOSCOPE METAL SHEATH, 24 FR; ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL Back to Search Results
Model Number EMRS-24
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A user facility reported that the distal tip of a scope was broken.It is unknown when the reported issue occurred.There was no patient injury or adverse event reported to olympus.
 
Additional Manufacturer Narrative
This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.The device was returned to olympus for evaluation where service confirmed the customer's complaint.During the evaluation, the complaint was confirmed.It was found that the insulating tip was chipped and damaged.In addition, meatus washer was broken off and damaged, and the red dot on the sheath head was missing.A review of the device history record found no deviations that could have caused or contributed to the reported issue.A definitive root cause cannot be identified.This information is addressed in the instructions for use (ifu): ¿warning: always keep sheaths parallel to one another when assembling and disassembling.If the inner sheath is inserted or removed at an angle to the outer sheath, the lateral force applied to the inner sheath may crack, loosen, break, or otherwise damage the sheath¿s insulated distal tip.A broken tip, or fragments of a damaged tip, can potentially pass through the outer sheath and into the patient.If drag or resistance is encountered during assembly or disassembly, stop¿align working element and sheath parallel to one another before proceeding.¿ olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NON-CF RESECTOSCOPE METAL SHEATH, 24 FR
Common Device Name
ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL
Manufacturer (Section D)
GYRUS ACMI, INC.
93 north pleasant st.
norwalk OH 44857
Manufacturer (Section G)
GYRUS ACMI INC.
800 west park drive
westborough MA 01581
Manufacturer Contact
todd brill
93 north pleasant st.
norwalk, OH 44857
5082077661
MDR Report Key15123177
Report Number1519132-2022-00012
Device Sequence Number7381949
Product Code FAS
UDI-Device Identifier00821925001961
UDI-Public00821925001961
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K890328
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEMRS-24
Device Lot NumberHW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2021
Initial Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2006
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-