Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 the mayfield triad skull clamp (a1108) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation showed that the unit passed all specific functional testing requirements, expect for the lock having rotational movement when the unit is not under pressure; however, this would not have caused a slippage.When the unit is properly positioned and put under pressure, the unit would not have slipped.Unrelated to the complaint issue, the unit was machined to have heli-coils added to the large starburst threads; unit was received without an adjustment wrench.General maintenance and cleaning performed.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.Repairs could not duplicate slippage.Probable root cause is improper or suboptimal positioning of the skull clamp.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|