MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SONICBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP; INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SONICBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP; INSTRUMENT Back to Search Results
Model Number SB-0545FC
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is returned and an evaluation completed for it.The user¿s complaint was confirmed.Upon inspection and testing, it was observed that the tissue pad was heavily worn.The probe had a crack at 12 mm from the tip.The probe had contact marks with the gripper.There were marks on the gripping area where the probe had come into contact with the gripper.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Event or Problem Description
As reported for this event by the customer, 30 minutes into a therapeutic laparoscopic cholecystectomy procedure a metallic noise was heard and the device tissue pad was found to be very worn.Error message receive is not known.The procedure was completed with a new similar device.There is no harm or adverse impact to the patient.The device was inspected prior to use with no anomaly noted.The setting at the time of the event were maximum 3/variable 1.Intelligent tissue monitoring (itm) functionality was on during the procedure.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.Ultrasonic output was performed while the tissue was gripped by the tip of the grip.2.The probe and tissue pad came into contact with each other on the rear end side of the grip, and the tissue pad was severely worn.3.Since the output was made with the probe and non-insulated part in contact, a metallic sound was generated.In addition, contact marks (scratches) were generated.4.Ultrasonic vibration and a load to grip the tissue are applied to the probe, causing cracks starting from scratches and causing an error.This following information is included in the instructions for use (ifu): ¿do not activate output while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ olympus will continue to monitor the performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP
Common Device Name
INSTRUMENT
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15125333
Report Number9614641-2022-00060
Device Sequence Number5736802
Product Code LFL
UDI-Device Identifier04953170370571
UDI-Public04953170370571
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSB-0545FC
Device Lot Number1YK 08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2022
Supplement Date Received by Manufacturer09/05/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
USG-400TD-SB400
Patient SexUnknown
-
-