• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LPS CALCAR PLANER SM; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LPS CALCAR PLANER SM; HIP INSTRUMENTS : REAMERS Back to Search Results
Model Number 2987-70-001
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the calcar reamer was dull.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a1, a2 (age), b5 and h6 (impact code).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that the surgery was delayed only minutes until the second set was opened.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4).Investigation summary
=
> the device associated with this report was received for examination.The investigation found the device dull to the touch, this is to be expected in a device that has been in the field for 12+ years.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Drawing/specifications reviewed? n/a the following document was reviewed: wi #103276989 | rev:3 device history lot
=
> a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the required lot code is a date code which not represents a finished goods lot#, however as it was determined that the reusable instrument device was worn from repeated use and servicing; further investigation for the reported complaint device is not required.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary - no device associated with this report was received for examination.The investigation could not verify or draw any conclusions about the root cause of the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot - a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS CALCAR PLANER SM
Common Device Name
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15126065
Report Number1818910-2022-14380
Device Sequence Number20665116
Product Code LYS
UDI-Device Identifier10603295158790
UDI-Public10603295158790
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2987-70-001
Device Catalogue Number298770001
Device Lot NumberHT0710
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer08/01/2022
08/23/2022
09/07/2022
09/09/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date08/03/2022
08/24/2022
09/08/2022
09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age50 YR
Patient SexUnknown
-
-