| Model Number |
SD-400U-10 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported to olympus, the facility had four (4) patients with post-polpectomy bleeding requiring treatment.Patient one (1) had a polpectomy for a pedunculated polyp of the rectosigmoid hinge.The patient had an immediate post-polypectomy hemorrhage.The treatment was clips.This event is captured in (b)(6).Patient two (2) had a polpectomy for a pedicled left colon polyp, herb scalpel.The patient had an immediate post-polypectomy hemorrhage.The treatment was clips.This event is captured in (b)(6).Patient three (3) had a polpectomy for a centimetric pedicle polyp of the sigmoid colon.The patient had an immediate post-polypectomy hemorrhage.The treatment was clips.This event is captured in (b)(6).Patient four (4) had a polpectomy for a pedunculated sigmoid polyp.The patient had an immediate post-polypectomy hemorrhage.The treatment was clips.The patient also had a late of bleeding which required hospitalization, deglobulation and a new endoscopic examination.This event is captured in (b)(6).The customer reported four (4) different patient events as follows: (b)(6): patient 1 - sd-400u-10, 21v; (b)(6): patient 2 - sd-400u-10, 24v; (b)(6): patient 3 - sd-400u-10, 24v; (b)(6): patient 4 - sd-400u-10, 24v.This is report 3 of 4 for (b)(6): patient 3 - sd-400u-10, 24v.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has not been returned to olympus for evaluation.The investigation is in process.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provided additional information from the customer.
|
| |
|
Event or Problem Description
|
|
Additional information provided by customer.The clips used for treatment were non-olympus devices (nova clip r2).
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide a correction and the results of the legal manufacturer¿s investigation and the device history record (dhr) review.New information was added to the following fields: the subject device was manufactured in april 2022 based on the provided 3 digit lot information.The dhr for this device was reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to resection of the polyp.There were no reports of device malfunctions and the device was not returned, therefore a causal relationship between the subject device and the bleeding could not be determined.The instructions for use (ifu) states the following: ·the operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique.This manual, therefore, does not explain or discuss clinical endoscopic procedures.·pull the slider to confirm that the distal end of the snare loop is completely retracted into the snare tube.When using the instrument while the snare loop is not completely retracted into the snare tube, the strangulation becomes insufficient and the tissue may not be resected completely and hemorrhages, punctures, or mucous membrane damage could result.In addition, it may be impossible to remove the snare loop from tissue due to burn on tissue.In this case, follow the instructions on page 12.·do not force the distal end of the insertion portion against body cavity tissue.This could cause patient injury, such as punctures, hemorrhages, or mucous membrane damage.·do not push the slider abruptly.The snare loop may extend abruptly.This could cause patient injury, such as punctures, hemorrhages, or mucous membrane damage.·do not use excessive force when snaring tissue.This could cause patient injury, such as hemorrhages or mucous membrane damage.·do not snare tissue with excessive force.This could break the snare loop.Olympus will continue to monitor the field performance of the device.
|
| |
|
Search Alerts/Recalls
|