Catalog Number 955792 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 07/05/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during continuous renal replacement therapy (crrt) on a prismaflex machine, a "scale error" alarm was generated and the screen froze and the clinician could not return the extracorporeal blood to the patient.A blood transfusion was performed.No additional information is available.
|
|
Manufacturer Narrative
|
The device was not received for evaluation.A device history review revealed no issues that could have caused or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
|