• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE; PKL, LIGATOR, HEMORRHOIDAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE; PKL, LIGATOR, HEMORRHOIDAL Back to Search Results
Model Number G58010
Device Problem Migration (4003)
Patient Problem Laceration(s) (1946)
Event Date 06/30/2022
Event Type  malfunction  
Event Description
During an esophagogastroduodenoscopy (egd), the physician used a cook instinct plus endoscopic clipping device.It was reported that the clip seemed to deploy properly but fell off seconds after deployment, leaving a slight amount of tissue damage.The procedure was completed with another clip.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Pma/510(k): k212323.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all instinct plus endoscopic clipping devices are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Type of Device
PKL, LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key15126627
MDR Text Key304461512
Report Number1037905-2022-00421
Device Sequence Number1
Product Code PKL
UDI-Device Identifier00827002580107
UDI-Public(01)00827002580107(17)250322(10)W4577511
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG58010
Device Catalogue NumberINSC-P-7-230-S
Device Lot NumberW4577511
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS 190 ENDOSCOPE
-
-