| Model Number |
4585 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Per medwatch received: battery component of wrist band forcefully separated or exploded while in use on patient.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.Visual inspection found the battery housing was separated and one battery was opened.The sensor passed continuity testing of the emitter and detector sites; however was unable to test device functionality due to the damaged sensor.Battery voltage testing could not be performed due to one opened battery and a non-functional battery pack.
|
| |
|