STRYKER ENDOSCOPY-SAN JOSE 1488 HD 3-CHIP CAMERA HEAD, C-MOUNT - INTERNATIONAL KIT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 1488210105
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
The device was received at the local service facility for investigation.The technical service report is attached (see communication log), and indicates: summary of evaluation: weak spot at the cable assembly insulation, scratches at the front enclosure window replaced the cable assembly, cmos kel cable, front enclosure and o-ring, completed qip.Camera head will be returned to the customer.Probable root cause: cables, connectors, digital board, power supply, filter / fuse, ac inlet board, main board, transition board, fiber board, intermittence between transition board and ch connector, software, scope, light source, camera head, coupler , 1488 & 1588 extension cable, intermittence while using rf devices (ex: valleylab), problem toggling light source or from camera head, connected monitors, leaking, use error, poor grounding, electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power, surge, manufacturing/ service nonconformity.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1488 HD 3-CHIP CAMERA HEAD, C-MOUNT - INTERNATIONAL KIT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15126937
Report Number0002936485-2022-00440
Device Sequence Number20484115
Product Code GCJ
UDI-Device Identifier07613327061055
UDI-Public07613327061055
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K132785
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1488210105
Device Catalogue Number1488210105I
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-