Model Number IPN920043
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "sheath was inserted.Attempt to advance a jl 6 guiding catheter failed, strong resistance was felt.Sheath and catheter were removed and a new sheath of the same lot was inserted.The same problem occurred, again the catheter could not be advanced.User changed to a terumo sheath which was successful." no patient harm reported.The patient's condition is reported as fine.It was reported the guide catheters that are inserted into the sheath have curved tips.When inserting the catheter through the sheath, resistance is felt near the tip of the sheath.The sheath seems to tighten in the inner lumen at the tip.
 
Event or Problem Description
It was reported that "sheath was inserted.Attempt to advance a jl 6 guiding catheter failed, strong resistance was felt.Sheath and catheter were removed and a new sheath of the same lot was inserted.The same problem occurred, again the catheter could not be advanced.User changed to a terumo sheath which was successful." no patient harm reported.The patient's condition is reported as fine.It was reported the guide catheters that are inserted into the sheath have curved tips.When inserting the catheter through the sheath, resistance is felt near the tip of the sheath.The sheath seems to tighten in the inner lumen at the tip.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CATH-LAB SHEATH INTRO SET:
Common Device Name
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15127105
Report Number9680794-2022-00463
Device Sequence Number17798622
Product Code DQY
UDI-Device Identifier10801902192445
UDI-Public10801902192445
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K862056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN920043
Device Catalogue NumberCL-07645
Device Lot Number14F21K0202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VSI CATHETER, CATALOG# 5571; VSI CATHETER, CATALOG# 5571
Patient SexUnknown
-
-