• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AUTOCAT 2; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AUTOCAT 2; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0400
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "during transport, it was found that the machine battery did not supply power.After the patient is transferred to the ward, iabp uses ac power to continue auxiliary support.There was no report of patient complications, serious injury or death".Patient condition reported as "fine".
 
Event or Problem Description
It was reported that "during transport, it was found that the machine battery did not supply power.After the patient is transferred to the ward, iabp uses ac power to continue auxiliary support.There was no report of patient complications, serious injury or death".Patient condition reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).The reported complaint of "machine battery did not supply power" is not able to be confirmed.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
 
Event or Problem Description
It was reported that "during transport, it was found that the machine battery did not supply power.After the patient is transferred to the ward, iabp uses ac power to continue auxiliary support.There was no report of patient complications, serious injury or death".Patient condition reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of "machine battery did not supply power" is not able to be confirmed.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.Other remarks: corrected data: additional information received: patient age, gender, and weight was added to section a.It was reported that "the patient's hemodynamics were unstable and blood pressure fluctuated".Upon arrival "at the ward, the equipment was connected to ac power and started to work, the patient's hemodynamics were stable, no medical intervention required".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT 2
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key15127110
Report Number3010532612-2022-00274
Device Sequence Number6603648
Product Code DSP
UDI-Device Identifier00801902051721
UDI-Public00801902051721
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K060309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberIAP-0400
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer08/19/2022
08/19/2022
Initial Report FDA Received Date07/28/2022
Supplement Report FDA Received Date08/26/2022
09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED; NONE REPORTED
Patient Age52 YR
Patient SexMale
Patient Weight75 KG
-
-