| Model Number |
401184 |
| Medical Device Problem Codes |
Failure to Capture (1081); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should the lead become available for analysis.
|
| |
|
Event or Problem Description
|
|
Lead dislodgement was reported.At the post implant follow-up, the lv lead did not capture in any configuration.Rv lead parameters were regular and stable.At the x ray the lv lead tip was in atrium.The patient does not appear to have done anything that may have caused the dislodgment.Planning lv lead revision.
|
| |
|