H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one vaccess pta catheter was returned for evaluation.No specific anomalies were noted during the visual evaluation.In functional testing, the balloon was inflated with an in-house presto device and the balloon was noted to be leaking.Under microscopic observation, a compound ruptured was noted.No other functional testing was performed.Therefore, the investigation is confirmed for the reported balloon rupture as a compound rupture was noted during the microscopic observation.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 04/2025), h11: h6(result, conclusion), h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|