COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G44154
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation = ir tech.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, during an "angio" of the right leg, the tip of a flexor ansel guiding sheath tore.Resistance was reportedly encountered upon advancement of the device over an unspecified 0.035 stiff angled glidewire, and the device reportedly bent.Upon removal of the device, it was noted that the tip was torn.The anatomy was mildly calcified.Another sheath was used to successfully complete the procedure without issue.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Additional information has been requested.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information was received 26jul2022 and was inadvertently omitted from the initial report.The hydrophilic coating had been activated.Access was obtained in the femoral artery.The sheath had not been advanced far when resistance was encountered and the device would no longer advance.Per the customer, the sheath and dilator tips were damaged and torn.
 
Event or Problem Description
Upon return and evaluation of the device on 23aug2022, the sheath tip was torn.No damage was found on the dilator.
 
Additional Manufacturer Narrative
Additional/corrected information: b5, h6 (annex g): only the sheath tip was split.No damage was noted to the dilator.H3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary: as reported, during an "angio" of the right leg, the tip of a flexor ansel guiding sheath tore.The hydrophilic coating had been activated.Access was obtained in the femoral artery.The sheath had not been advanced far when resistance was encountered.The device would no longer advance over an unspecified 0.035 stiff angled glidewire, and the device reportedly bent.Upon removal of the device, it was noted that the tip was torn.The anatomy was mildly calcified.Another sheath was used to successfully complete the procedure without issue.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures and a visual inspection of the device were conducted during the investigation.One kcfw-6.0-18/38-45-rb-anl0-hc was returned for investigation.The distal tip of the sheath was buckled and split.A document-based investigation evaluation was performed.No related non-conformances were recorded, and there have been no other reported complaints for this lot number.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which warn, ¿if resistance is encountered during advancement of the flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.¿ the ifu further states, ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ based on the available information, cook has concluded that a component failure contributed to this failure mode.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15128397
Report Number1820334-2022-01295
Device Sequence Number5667022
Product Code DYB
UDI-Device Identifier00827002441545
UDI-Public(01)00827002441545(17)250331(10)14639822
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model NumberG44154
Device Catalogue NumberKCFW-6.0-18/38-45-RB-ANL0-HC
Device Lot Number14639822
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/19/2022
Supplement Date Received by Manufacturer07/26/2022
08/23/2022
10/21/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date08/09/2022
09/16/2022
11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-