Model Number 407451
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2022
Event Type  malfunction  
Event Description
During an ep study, a blood leak was noted in the sheath.Another device was used to complete the procedure with no consequences to the patient.
 
Manufacturer Narrative
One 8.5f swartz braided introducer sheath was received for evaluation.Functional testing confirmed a fluid leak in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak remains unknown.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0¿, 63 CM LENGTH, 8.5 F
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICALL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15128770
MDR Text Key302427641
Report Number3005334138-2022-00438
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734205757
UDI-Public05414734205757
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number407451
Device Catalogue Number407451
Device Lot Number8491180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-