ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0¿, 63 CM LENGTH, 8.5 F; INTRODUCER, CATHETER
|
Back to Search Results |
|
Model Number 407451 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/15/2022 |
Event Type
malfunction
|
Event Description
|
During an ep study, a blood leak was noted in the sheath.Another device was used to complete the procedure with no consequences to the patient.
|
|
Manufacturer Narrative
|
One 8.5f swartz braided introducer sheath was received for evaluation.Functional testing confirmed a fluid leak in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak remains unknown.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
|
|