• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN EXTRACTION BALLOON V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN EXTRACTION BALLOON V Back to Search Results
Model Number B-V243Q-A
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reports during an endoscopic retrograde cholangiopancreatography (ercp) using an evis exera iii duodenoscope with a single use distal cover and a single use extraction balloon, the distal cover split open during the procedure and several devices broke because of it.It was observed that bioburden had gotten lodged behind the elevator of the scope.The elevator could not fully function in its range of motion, could not go fully up and down.At first, the customer noticed that the device (scope/distal cover) was not able to pass smoothly through the channel, and the physician observed that he was not able to fully open or close the elevator.The elevator jammed at 45 degrees angle and was not able to fully open.When passing devices, the physician ended up stripping and breaking some of the devices.An extraction balloon that was broken up inside the patient's duct and they had to take scope out and go retrieve it.Once retrieved, the physician to use a different balloon and it started to break because could not fully open elevator, it felt jammed.The physician placed a stent.After barely placing the stent, they finished the procedure.The physician took scope out and realized cap was completely split at the top, and there was a lot of bio burden back behind the elevator causing it to not fully open and close.There were no procedural or anatomical challenges that could have contributed to the distal cover splitting.The lot number cannot be provided for the distal cover used.The scope is not being returned to olympus for repair.Once the scope was cleaned, the scope was found to be in working order.The elevator was moving without issue.The physician was successful in sweeping the stones, but the balloon popped due to sharp stone edges.Another balloon ripped before inserting it into duct.In hindsight, the elevator was not moving appropriately due to sludge and stone jamming behind the elevator channel.This was not known at the time of the procedure.A non-olympus retrieval device was used to extract the balloon from the patient¿s duct.It was observed that the balloon was stretching inappropriately.There were no adverse effects to the patient as a result of this occurrence.The procedure was completed.The patient¿s current condition is reported as discharged home in stable condition.Case with patient identifier (b)(6) reports the b-v243q-a.Case with patient identifier (b)(6) reports the tjf-q190v.Case with patient identifier (b)(6) reports the maj-2315.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN EXTRACTION BALLOON V
Common Device Name
SINGLE USE 3-LUMEN EXTRACTION BALLOON
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
MDR Report Key15128789
Report Number2429304-2022-00010
Device Sequence Number6604204
Product Code LQR
UDI-Device Identifier04953170368721
UDI-Public04953170368721
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/29/2022
Report Date (Section F) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB-V243Q-A
Device Lot Number22V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/30/2022
Event Location Hospital
Date Report to Manufacturer06/30/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/29/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TJF-Q190V, SN (B)(4), MAJ-2315, LOT UNK
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age71 YR
Patient SexFemale
Patient Weight76 KG
-
-