| Model Number |
37800 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Erosion (1750); Inflammation (1932); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2020 product type lead.Product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2020 product type lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 04-nov-2021, udi#: (b)(4); product id: 4351-35, serial/lot #: (b)(4), ubd: 16-apr-2022, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implan ted with an implantable neurostimulator (ins) for gastric stimulation and gastrointestinal/ pelvic floor.It was reported that patient showed up to the er at (b)(6) in (b)(6), tn with abdominal pain.They ended up doing a colon resection as the leads had adhered to her colon.Cause was reported to be lead adhesion to colon.Hcp was able to remove the adhesion and free the leads up within the abdomen on (b)(6) 2022.The issue was confirmed resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id 4351-35 lot# serial# (b)(6) implanted: (b)(6)2020.Explanted: product type lead product id 4351-35 lot# serial# (b)(6).Implanted: (b)(6)2020.Explanted: product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received from a health care provider (hcp) via a manufacturer representative(rep).It was stated that the patient reported to an er in morristown, tn in july 2022 with abdominal pain and inflammation.Ct scan was abnormal.They took the patient to the or and found that their bowel had wrapped around the leads.They performed a hemicolectomy.Patient was seen by the surgeon who performed the surgery on 8/8/22 for follow-upand they were doing well.Patient has a history of fibromyalgia.Enterra battery and leads are performing well and impedance is in range.
|
| |
|
Search Alerts/Recalls
|