| Model Number |
380320 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The passeo-14 balloon catheter was selected for treatment of a mildly calcified lesion with 80 percent stenosis degree in the posterior tibial artery.The passeo-14 was advanced over the wire and inflated at 12 bar.The balloon couldn't inflate further and leaking of contrast was observed in the fluoroscopy.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed a fine jet of water emerging from the distal part of the balloon when attempting inflation.Microscopic analysis of the balloon surface showed a small pinhole direct at the distal edge of the distal x-ray marker.In close vicinity of the pinhole scratches and shatter marks were observed which have likely been caused by a hard, sharp-edged object such as e.G.Anatomical structure.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test and a pressure test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause for the reported event is most likely related to the patients anatomy.
|
| |
|
Search Alerts/Recalls
|