• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number PXT261414
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Aneurysm (1708); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/27/2012
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).It should be noted the gore® excluder® aaa endoprosthesis instructions for use (ifu) state ¿adverse events that may occur and / or require intervention include, but are not limited to, endoleak.¿ per ifu, users are made aware of the risks associated with type ii endoleaks and are instructed to consider the risks and benefits discussed in the ifu for each patient before using the devices.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2012.The patient underwent endovascular treatment of an abdominal aortic aneurysm using a gore® excluder® aaa endoprosthesis.During the procedure, a type ii endoleak was reportedly observed.The physician elected to monitor the endoleak, and the procedure was concluded without further treatment.On an unknown date in 2016, aneurysm enlargement was reportedly observed, and the patient underwent open surgical repair.According to the report, aneurysmorrhaphy and branch ligation including the lumbar artery were performed.The type ii endoleak was resolved, and the patient tolerated the procedure.
 
Additional Manufacturer Narrative
H6: code c19 ¿ a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: conclusion coding updated.It should be noted the gore® excluder® aaa endoprosthesis instructions for use (ifu) state ¿adverse events that may occur and / or require intervention include, but are not limited to, endoleak.¿ per ifu, users are made aware of the risks associated with type ii endoleaks and are instructed to consider the risks and benefits discussed in the ifu for each patient before using the devices.H6: code e2402 used to capture type ii endoleak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY MOUNTAIN B/P
3750 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
damon jackson
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15130127
Report Number2017233-2022-03172
Device Sequence Number20489336
Product Code MIH
UDI-Device Identifier00733132610228
UDI-Public00733132610228
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Remedial Action Patient Monitoring
Type of Report Initial,Followup
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2013
Device Model NumberPXT261414
Device Catalogue NumberPXT261414
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer08/29/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date08/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/24/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age85 YR
Patient SexMale
-
-