• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. CLASSIC SCREWDRIVER; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. CLASSIC SCREWDRIVER; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3558-1
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event or Problem Description
It was reported that a crack was found in the end of a screwdriver during an inspection.No patient was present at the time the crack was found.
 
Additional Manufacturer Narrative
Corrections in h3.Additional information in h6: component, investigation type, findings, and conclusions.Inspection: visual inspection revealed a crack in the tip of the screwdriver.Dhr review the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis forces applied during use.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Event or Problem Description
It was reported that a crack was found in the end of a screwdriver during an inspection.No patient was present at the time the crack was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLASSIC SCREWDRIVER
Common Device Name
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15130257
Report Number3012447612-2022-00182
Device Sequence Number20453563
Product Code NKB
UDI-Device Identifier00889024336377
UDI-Public(01)00889024336377(10)68NR
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K132884
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3558-1
Device Lot Number68NR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Supplement Date Received by Manufacturer01/17/2023
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-