Corrected udi number.The affected pk papyrus stent system was returned to biotronik and subjected to a detailed technical investigation.Images of the case such as angiographies could not be obtained.The technical investigation revealed that the hypotube has fractured about 158 mm distal to the kink protector.At the fracture sites the crosssection of the hypotube is no longer circular but compressed into an ellipse and the polymer coating is plastically deformed.In addition, the hypotube was found kinked close to the fracture sites.The findings indicate that the hypotube most probably fractured as a result of significant bending outside of the patients body.In addition, the stent was found displaced on the balloon by about 8 mm in distal direction.The provided clinical information (pk papyrus device registration form, cathlab report) was reviewed to establish whether a device deficiency contributed to this event.Review of the product release documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the conducted investigations and the review of the provided documentation, no manufacturing related root cause could be identified.The root cause for the complaint event is most likely related to the handling of the device.Please note that the ifu advises the user to exercise care during device handling to reduce the possibility of accidental breakage, bending or kinking of the stent system shaft.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
|