| Model Number |
24706 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The patient presented with peripheral arterial disease and underwent endovascular therapy.The 99% stenosed target lesion was located in the moderately tortuous and mildly calcified below-the-knee vessel.A 3.0mm x 220mm x 150cm coyote balloon catheter was advanced for dilatation.However, during the first inflation at 14 atmospheres for 30 seconds, the balloon ruptured.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient was in good condition post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a longitudinal tear 25mm from the tip and is approximately 11mm long.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The patient presented with peripheral arterial disease and underwent endovascular therapy.The 99% stenosed target lesion was located in the moderately tortuous and mildly calcified below-the-knee vessel.A 3.0mm x 220mm x 150cm coyote balloon catheter was advanced for dilatation.However, during the first inflation at 14 atmospheres for 30 seconds, the balloon ruptured.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient was in good condition post procedure.
|
| |
|
Search Alerts/Recalls
|