Model Number TRX-615
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 12/10/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Patient reported "right-side exchange from textured to smooth breast implant due to the patient¿s concern with the product".Healthcare additionally reported upon removal capsular contracture, baker grade unknown.Device has been explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified; fold creases, wear abrasion, yellow particles, weight within specification.After autoclave cycle was identified: cloudy gel and bubbles in gel.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15131695
Report Number9617229-2021-60233
Device Sequence Number17559487
Product Code FTR
UDI-Device Identifier10888628008595
UDI-Public10888628008595
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/24/2021
Device Model NumberTRX-615
Device Catalogue NumberTRX-615
Device Lot Number2875721
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2021
Initial Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-