| Model Number |
26012804 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377); Metal Related Pathology (4530)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This event will be updated once the investigation is complete.Trends will be evaluated.
|
| |
|
Event or Problem Description
|
|
Allegedly, patient has been experiencing pain from around 9 months post-op and blood tests showed raised metal ion levels.Revision was performed to remove the original stem and 28mm metal skirted head.There is metal-on-metal wear on the trunnion.
|
| |
|
Event or Problem Description
|
|
Allegedly, patient has been experiencing pain from around 9 months post-op and blood tests showed raised metal ion levels.Revision was performed due to lysis, to remove the original stem and 28mm metal skirted head.There is metal on-metal wear on the trunnion.Revision njr number: (b)(4).Side: r primary asa: p2 - mild disease not incapacitating.Non-revised products: product id: pha06254 acetabular cup "procotyl® l" beaded coated+hap 44 grp b/ lot number: 1663283/ qty: 1.Product id: pha04604 rim-lock "a-class®" acetabular poly liner 0° size 28 group b/ lot number: 385786881650139/ qty: 1.
|
| |
|
Additional Manufacturer Narrative
|
|
Section a.1: additional information in patient / section b.5: additional information in description/ section h.6: health code added.
|
| |
|