• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 2.4MM TI VA LOCKING SCREW STARDRIVE 16MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.4MM TI VA LOCKING SCREW STARDRIVE 16MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.210.116
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in france as follows: the patient was hospitalized (b)(6) 2022.For a fracture of the distal radius reduced by osteosynthesis.The exact implant date is unknown.The patient was re-examined after one month and presented with unscrewing of the implants placed as well as breakage of the 4 distal epiphyseal locking screws at the body-head junction in the absence of a fall.Additionally secondary displacement of the fracture site with extra-articular callus occurred.The material initially implanted was removed during a surgery revision on (b)(6) 2022 to rectify the callus and new osteosynthesis by plate.This is report 1 of 9 for (b)(4).This report is for a locking screw.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.No code available is for injury/revision surgery.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: d9, h3, h4: a product investigation was completed: visual analysis of the returned sample revealed that the head of the locking screw broke off.The broken fragment was returned attached to the plate.No other problems identified.A dimensional inspection was unable to be performed due to post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.The current drawing was reviewed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary : product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.4MM TI VA LOCKING SCREW STARDRIVE 16MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15131758
Report Number8030965-2022-05414
Device Sequence Number5766494
Product Code HWC
UDI-Device Identifier10886982118523
UDI-Public(01)10886982118523
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K103243
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.210.116
Device Catalogue Number04.210.116
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/04/2022
Supplement Date Received by Manufacturer08/29/2022
09/05/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date08/29/2022
09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORTSCR Ø2.7 SELF-TAP L16 TAN; CORTSCR Ø2.7 SELF-TAP L16 TAN; CORTSCR Ø2.7 SELF-TAP L16 TAN; CORTSCR Ø2.7 SELF-TAP L16 TAN; CORTSCR Ø2.7 SELF-TAP L16 TAN; VA LOCKSCR Ø2.4 SELF-TAP L16 TAN; VA LOCKSCR Ø2.4 SELF-TAP L18 TAN; VA LOCKSCR Ø2.4 SELF-TAP L20 TAN; VA-LCP-2-COLUMN DRP2.4 VOLAR R SHAFT 2HO
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-