The patient's previous driveline communication fault is reported under mfr #: 2916596-2022-11218.Manufacturer's investigation conclusion: the reported event of the patient experiencing a driveline communication fault was confirmed via analysis of the submitted log file; however, the reported event could not be reproduced during testing.The heartmate 3 vad modular cable was returned and evaluated at abbott.During the evaluation, the cable was observed and there was some corrosion buildup inside the inline connector next to the pins.The modular cable was functionally tested and passed all steps without issue.The modular cable was placed on a mock circulatory loop for an extended period of time and was able to support pump function without any atypical alarms or issues active.The cable was manipulated in an attempt to reproduce the reported event; however, the cable functioned as intended.The modular cable was then dissected to examine the underlying wires of the cable to reveal that the yellow (comm b) wire contained a slight cut in the insulation, exposing the underlying conductor.This could have caused a short, causing the controller to alarm for a driveline communication fault; however, the root cause is unknown due to the modular cable not being able to reproduce the reported event.Additionally, the submitted log file (labeled 125036) contained data spanning approximately 18 hours (b)(6) 2022 ¿ (b)(6) 2022).The log file captured a driveline communication fault due to a com a fault active on (b)(6) 2022 from 8:35:29 to 8:46:15.The alarm did not resolve on it own.The driveline was immediately disconnected in the next event and the controller was turned off.Pump speed was not affected by the alarm.The root cause of the reported event could not be conclusively determined through this analysis; however, the corrosion inside the inline connector or the exposed conductor in the underlying wire of the modular cable could have contributed to the event.Incidental findings: corrosion inside the inline connector next to the pins, kinked underlying wires with exposed conductor the device history records were reviewed and the records revealed the heartmate 3 vad modular cable (lot#: 7132509) was manufactured in accordance with manufacturing and quality assurance specifications.The heartmate iii instructions for use (rev.G) section 7 entitled ¿alarms and troubleshooting¿ and heartmate iii patient handbook (rev.G) entitled ¿alarms and troubleshooting¿ addresses how to properly interpret and troubleshoot all system alarms including driveline fault alarm conditions, and the actions to take if the alarms cannot be resolved.The heartmate iii patient handbook (rev.G) "living with the heartmate 3" (under "caring for the driveline") contains information regarding how to clean, care for the driveline, and connect it properly.This section instructs the user to inspect the modular cable in-line connector for signs of damage, such as cracking, fraying, wear, exposed wires, sharp bends, or kinks.Call your hospital contact right away if the driveline is damaged (or might be damaged)." the heartmate iii patient handbook (rev.G)¿safety checklists¿ provides checklists to assist the patient in performing routine maintenance of heartmate iii lvad, including inspecting the driveline cable for signs of damage.This section also informs the user to replace any equipment or system component that appears damaged or worn.The heartmate iii patient handbook (rev.G) and the instructions for use (rev.G) cautions the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|