| Model Number |
186770445 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: kwq and kwp.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in australia as follows: it was reported that on (b)(6) 2022, a patient underwent a revision surgery to replace two viper prime screws that had broken at the distal tips.The surgeon used trephines and a screw extraction device to remove and replace with larger screws.The procedure was successfully completed with the use a of a competitor¿s product.This report involves one (1) viper prime cfx fen x-tab polyaxial screw 5.5 7 x 45mm.This is report 3 of 4 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: a review of the receiving inspection (ri) for viper prime cfxfen xtab 7x45mm was conducted identifying that lot number tbwgs was released in a single batch.Batch1: lot qty of (b)(4) units were released on aug 27, 2018 with no discrepancies.Supplier: tomz corp.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the distal tip of the viper prime cfxfen xtab 7x45mm p/n:186770445 was broken.The broken fragment was returned.No other problems identified.A dimensional inspection was unable to be performed due to post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper prime cfxfen xtab 7x45mm p/n:186770445 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|