• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. 4K UHD LCD MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. 4K UHD LCD MONITOR Back to Search Results
Model Number OEV321UH
Device Problems Loss of Power (1475); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2022
Event Type  malfunction  
Event Description
The customer reported to olympus, during an endoscopy procedure, the 4k uhd lcd monitor turned off and could not turn back on.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
The subject device has not been returned to olympus for evaluation.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
Additional information was received from the reporter indicating, the intended therapeutic procedure was completed with a similar device.It was also noted, extra sedation was required due to this event.The subject device has been returned to olympus for evaluation.During evaluation, it was observed, the monitor had no power, due to faulty power supply and it was recommended that the monitor be replaced.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to a faulty power supply.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4K UHD LCD MONITOR
Type of Device
MONITOR
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15132589
MDR Text Key304722101
Report Number3002808148-2022-00493
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170418723
UDI-Public04953170418723
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K220069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOEV321UH
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/03/2022
Initial Date FDA Received07/29/2022
Supplement Dates Manufacturer Received08/01/2022
09/13/2022
Supplement Dates FDA Received08/25/2022
10/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-