• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH 5.5 TI CORT FIX 6X45MM; ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH 5.5 TI CORT FIX 6X45MM; ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION Back to Search Results
Model Number 186731645
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional device product codes: kwp, mni, nkb, kwq.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2022, the surgeon was inserting rods for a ¿dry-fit¿ before cement injection and discovered that the saddle inside of the poly head of the screw had been broken.This would not allow the rod to seat in the poly head and would also make attachment of the cement cannulas impossible.There were no broken fragments.The screw was removed without issue and another was inserted.The procedure was completed successfully with no surgical delay reported.There was no patient consequences or outcomes reported.This report is for one (1) 5.5 ti cort fix 6x45mm.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: a manufacturing record evaluation was performed for the finished device product code: 186731645; lot number: 156729.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: september 13, 2017.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the 5.5 ti cort fix 6x45mm, p/n: 186731645.A dimensional inspection was unable to be performed due to device design.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the 5.5 ti cort fix 6x45mm, p/n: 186731645 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 TI CORT FIX 6X45MM
Common Device Name
ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
8472871282
MDR Report Key15132895
Report Number1526439-2022-01306
Device Sequence Number20451698
Product Code MNH
UDI-Device Identifier10705034351834
UDI-Public(01)10705034351834
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K110216
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186731645
Device Catalogue Number186731645
Device Lot Number156729
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer08/03/2022
Initial Report FDA Received Date07/29/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/13/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-