The initial reporter received questionable elecsys ft4 assay, probnp g2 elecsys e2g v2, elecsys ferritin, folate, elecsys tsh assay and elecsys vitamin b12 immunoassay results for two patient samples tested on the cobas pro sbl e 801.For sample 2063033611, the initial ft4 result was reported to the physician.The physician did not believe the result prompting the rerun of several assays.Discrepant results were observed for ft4 and probnp.On (b)(6) 2022, sample 1's initial ft4 result was 1.820 pmol/l with a data flag.The repeat result was 40.9 pmol/l.The initial probnp result was <1 pmol/l with a data flag.The repeat result was 67 pmol/l.On (b)(6) 2022, sample 2071221711's initial ferritin result was 1.640 ug/l with a data flag.The repeat result was 114 ug/l.The initial folate result was <2.72 nmol/l with a data flag.The repeat result was 45.100 nmol/l.The initial tsh result was <0.1 miu/l with a data flag.The repeat result was 5.000 miu/l.The initial vitamin b12 result was <73.8 pmol/l with a data flag.The repeat result was 366 pmol/l.The repeat results were deemed correct.The ft4 reagent lot number is 60647801 with an expiration date of 30-nov-2022.The probnp reagent lot number is 60018601 with an expiration date of 31-may-2022.The ferritin, folate, tsh and vitamin b12 reagent lot numbers and their expiration dates were requested but not provided.
|