|
W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
PXT261414 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).It should be noted the gore® excluder® aaa endoprosthesis instructions for use (ifu) state ¿adverse events that may occur and / or require intervention include, but are not limited to, endoleak.¿ w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2012, the patient underwent endovascular treatment of an abdominal aortic aneurysm using a gore® excluder® aaa endoprosthesis.During the procedure, a type ii endoleak was reportedly observed.The physician elected to monitor the endoleak, and the procedure was concluded without further treatment.On (b)(6) 2022, gradual enlargement of the aneurysm was observed.Abdominal ultrasonography reportedly revealed a proximal type i endoleak.It was suggested that type i appeared because the seal zone became shorter as the diameter of the aneurysm enlarged.On (b)(6) 2022, the patient underwent reintervention whereby additional stent grafts were implanted to extend the device system proximally, and to extend the bilateral limbs distally.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: code c19 ¿ a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: conclusion coding updated.It should be noted the gore® excluder® aaa endoprosthesis instructions for use (ifu) state ¿adverse events that may occur and / or require intervention include, but are not limited to, endoleak, and aneurysm enlargement.¿.
|
| |
|
Search Alerts/Recalls
|
|
|