| Model Number |
500-55112 |
| Medical Device Problem Codes |
Device Alarm System (1012); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device eval by manufacturer: the usc and the ic were returned.The usc had a pinch/kink at 10.3cm from the strain relief.Blood was seen in the ic drug lines.Five of the drug holes on the ic appeared occluded with blood.During water bath testing the ic was able to be flushed with air and the majority of the drug holes showed bubbles.An infusion pump was connected to the drug port to see if the blood occluded holes would cause a pump alarm.The pump alarmed after about 1minute of run time and was cleared and continued to alarm every 1-2minutes until it was shut off after 7minutes 30 seconds.
|
| |
|
Event or Problem Description
|
|
It was reported that a down stream drug occlusion occurred requiring additional intervention.Two 106cm x12cm ekosonic endovascular catheters were selected for treatment of a pulmonary embolism.The patient was brought from the interventional radiology (ir) lab to the intensive care unit (icu) with the catheters placed but without tpa infusing.When the catheters clotted the patient was taken back to the ir lab.One of the catheters was replaced.The second catheter was repositioned and the patient was returned to the icu.Upon the patient's return to the icu, therapy was restarted and running when an alarm indicating a downstream occlusion on both drug lines was received.The catheter was checked for kinks, closed stopcocks and closed roller clamps; there were none.Infusion rate and pump pressure were within the recommended range.The catheters were checked for coils or sutures; there were none.The nurse attempted to flush the catheters but was unsuccessful.The catheter was pulled back 1mm and flushing was reattempted but the line was unable to be flushed due to clotting.The patient was worsening possibly due to delay in treatment caused by the clotting of the catheters.The patient was placed on extracorporeal membrane oxygenation (ecmo) due to right ventricular failure and was transferred to another health care facility where she is under supervision and doing well.
|
| |
|
Search Alerts/Recalls
|