Model Number 7601-05026
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
This record is a consolidation of q2 2022 records summarized as a part of the fda voluntary malfunction summary reporting program.This report captures one instance of intraoperative yukon polyaxial screw fracture.The associated procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Additional Manufacturer Narrative
The screw was visually analyzed for macroscopic indications of failure mode.The fracture occurred immediately beneath the ball feature at the screw neck.The fracture surface was relatively level and the fracture plane was orthogonal to the screw axis.The fracture surface showed plastic deformation and helical distortion characteristic of ductile failure in torsion.The center of the helical distortion is slightly off-center with respect to the screw axis.The direction of the fracture dimples about the origin indicate the screw fractured during insertion.There is significant deformation on and material lost from the screw threads at the top of the shank, likely caused by the implement used to extract the screw after it fractured (i.E., pliers).Several chunks of the threads are missing near the distal point of the shank; this may have been caused during removal or during handling and transit after extraction.Device history records were reviewed and no relevant manufacturing issues were identified.Complaint history for the reported lot number was reviewed and no adverse trends were identified.The most likely root cause is stress beyond the structural capacity of the screw.Various factors may have contributed to the failure, including, but not limited to: screw diameter (larger screws require more torque to insert); application of off-axis forces; screw hole prep (under-tapping, not tapping, small drill size, shallow drilling depth, etc.); patient factors (bone quality, pathologies, etc.).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL SCREW; SIZE Ø5.0X26 MM
Common Device Name
POSTERIOR CERVICAL SCREW SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key15135739
Report Number3004774118-2022-00301
Device Sequence Number17451737
Product Code NKG
UDI-Device Identifier10888857350007
UDI-Public10888857350007
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K171444
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7601-05026
Device Catalogue Number7601-05026
Device Lot NumberJKCR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2022
Initial Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-